Potential inaccuracies in mass spectrometers with spectrum overlap erasure units used during anaesthesia.

作者信息

Beatty P C

出版信息

Clin Phys Physiol Meas. 1984 May;5(2):93-104. doi: 10.1088/0143-0815/5/2/005.

Abstract

Cracking pattern overlap effects have been identified as one of the major sources of measurement error in the use of mass spectrometers in anaesthesia. Spectrum overlap erasure units, which subtract from one signal peak a portion of another peak, have been used to eliminate overlap due to cracking pattern. Using a computer simulation, the efficacy of various spectrum overlap erasure configurations with associated automatic sensitivity controls has been studied. The effect of noise and changes in cracking pattern have been assessed. Accuracies predicted by the simulation have been compared with an arbitrary standard of accuracy (0.1% vol/vol), for a selection of gases commonly encountered in clinical anaesthesia.

摘要

裂解模式重叠效应已被确认为麻醉中使用质谱仪时测量误差的主要来源之一。光谱重叠消除单元从一个信号峰中减去另一个峰的一部分,已被用于消除由于裂解模式导致的重叠。通过计算机模拟,研究了各种具有相关自动灵敏度控制的光谱重叠消除配置的功效。评估了噪声和裂解模式变化的影响。对于临床麻醉中常见的几种气体,将模拟预测的准确度与任意准确度标准(0.1%体积/体积)进行了比较。