Beck J, Schettini A, Salton R
J Neurosurg. 1984 Oct;61(4):733-6. doi: 10.3171/jns.1984.61.4.0733.
The application of the coplanar principle to dynamic epidural pressure measurements was investigated in vitro. The authors used a coplanar pressure-displacement transducer, commonly employed to measure the viscoelastic properties of brain tissue in vivo. The present studies were performed using canine dura and a specially constructed fluid-filled chamber. The accuracy of the technique was assessed by comparing the pressure in the chamber recorded by the coplanar transducer to the pressure measured by a transducer directly vented to the chamber. The results show that the coplanar principle remained valid for dynamic measurements with the transducer under a variety of conditions.
在体外研究了共面原理在动态硬膜外压力测量中的应用。作者使用了一种共面压力-位移传感器,这种传感器常用于体内测量脑组织的粘弹性特性。本研究使用犬类硬脑膜和一个特制的充液腔室进行。通过比较共面传感器记录的腔室内压力与直接通向腔室的传感器测量的压力,评估了该技术的准确性。结果表明,在各种条件下,共面原理对于使用该传感器进行动态测量仍然有效。