Munson G, Satterlee C, Hammond S, Betten R, Gaines R W
Clin Orthop Relat Res. 1984 Oct(189):97-102.
Segments of the spinal column of calves were instrumented with Harrington rods with and without sublaminar wires. Grossly unstable lesions were created by resecting a 2-cm segment from the central spinal segment. Testing on an Instron tester documented the improvement in stability provided by the sublaminar wires in this grossly unstable experimental fracture. Improved stability was seen in resisting axial loading, lateral bending, and forward flexion loading. Rotational stability was particularly improved.
对小牛的脊柱节段使用哈林顿棒并分别带有和不带有椎板下钢丝进行器械固定。通过从脊柱中央节段切除2厘米的节段来制造严重不稳定的损伤。在英斯特朗试验机上进行的测试记录了在这种严重不稳定的实验性骨折中椎板下钢丝所提供的稳定性改善情况。在抵抗轴向载荷、侧弯和前屈载荷方面观察到稳定性有所提高。旋转稳定性尤其得到了改善。