The search of corrosion inhibitors reducing the corrosive action of the sterilizing and disinfecting media has been performed in order to protect instruments made of metals against corrosion during these processes. The program of the investigations includes potentiodynamic and potentiostatic measurements and full-scale tests. The infection of the sodium benzoate or potassium gluconate into the disinfecting chloramine solution and sterilizing hydrogene peroxide solution has been shown to improve the resistance to the corrosion for medical instruments made of carbon steel.
为了保护金属制成的器械在消毒和灭菌过程中不受腐蚀,人们一直在寻找能降低消毒和灭菌介质腐蚀作用的缓蚀剂。研究方案包括动电位和恒电位测量以及全面测试。已表明,将苯甲酸钠或葡萄糖酸钾加入消毒用的氯胺溶液和灭菌用的过氧化氢溶液中,可提高碳钢制成的医疗器械的耐腐蚀性。