Bogomolova S N, Gordashnik K Z, Gerasimov I V, Marinich M A
Med Tekh. 1984 Sep-Oct(5):45-8.
Considerations are given to some methodical problems, and corrosion performance of stainless steels 30 X 13 and 03X11H10 M2T2 used for production of medical instruments is studied by using the method of anode polarization curves. The treatment in hydrogen peroxide is shown to exert a distinct passivating effect on surfaces of the steels under study. Among the bioactive media the plasma substituting agent gelatinol is the most aggressive. When studying the corrosion resistance of metals by potentiostatic methods the physiological solution imitating the blood is proposed to be used as an operating medium.
考虑了一些方法性问题,并采用阳极极化曲线法研究了用于生产医疗器械的30X13和03X11H10M2T2不锈钢的腐蚀性能。结果表明,过氧化氢处理对所研究的钢表面具有明显的钝化作用。在生物活性介质中,血浆替代剂明胶醇的腐蚀性最强。在用恒电位法研究金属的耐腐蚀性时,建议使用模拟血液的生理溶液作为工作介质。