不同刚度特性的外固定装置下截骨愈合情况的比较。

Comparison of osteotomy healing under external fixation devices with different stiffness characteristics.

作者信息

Wu J J, Shyr H S, Chao E Y, Kelly P J

出版信息

J Bone Joint Surg Am. 1984 Oct;66(8):1258-64.

PMID:6490701
Abstract

The aim of this study was to investigate the effect of the rigidity of external fixation on osteotomy healing based on radiographic, morphological, histological, and biomechanical analyses of canine tibial osteotomies treated with more rigid (six half-pins) and less rigid (four half-pins) unilateral external fixators. Successful healing was obtained in all tibiae, but the bone-healing mechanisms appeared to be different depending on the rigidity of the device used. On the side with more rigid fixation, early clinical union was observed thirty days after osteotomy. Periosteal callus formation was significantly increased on the less rigidly fixed side at both ninety and 120 days (p less than 0.05). Fixation stiffness was found to affect the incidence of pin loosening (p less than 0.01) and of bone porosity (p less than 0.05) at the osteotomy site. However, new-bone formation and the structural strength of the bone were not significantly different between the four-pin and six-pin sides. These results seem to indicate that a longer period is required for fracture repair and remodeling when external fixation with less rigidity is used.

摘要

本研究旨在通过对采用更坚固(六根半针)和较不坚固(四根半针)单侧外固定器治疗的犬胫骨截骨术进行影像学、形态学、组织学和生物力学分析,探讨外固定刚度对截骨愈合的影响。所有胫骨均实现了成功愈合,但根据所用器械的刚度不同,骨愈合机制似乎有所差异。在固定更坚固的一侧,截骨术后30天观察到早期临床愈合。在固定较不坚固的一侧,90天和120天时骨膜骨痂形成均显著增加(p小于0.05)。发现固定刚度会影响截骨部位的钢针松动发生率(p小于0.01)和骨孔隙率(p小于0.05)。然而,四根针侧和六根针侧之间的新骨形成和骨的结构强度并无显著差异。这些结果似乎表明,使用刚度较低的外固定时,骨折修复和重塑需要更长时间。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验