[直线加速器面向实践的日常控制测量的可能性与局限性]

[Possibilities and limits of practice-oriented daily control measurements of linear accelerators].

作者信息

Kaulich T W, Körber S D

出版信息

Strahlentherapie. 1984 Sep;160(9):543-8.

PMID:6495357
Abstract

Daily control measurements of the dose taken at a linear accelerator are described in order to present a practice-oriented measurement and evaluation method rendering possible a long-term quality control. This method allows to establish sudden parameter variations as well as to determine in a simple and precise manner long-term drifting effects with an amplitude within the range of measurement variations and a period of many months. During the time of six months, not only a drifting effect with a long period and an amplitude within the range of pro mille, but also a variation coefficient smaller than 0,3% could be determined.

摘要

本文描述了线性加速器每日剂量控制测量方法,旨在提供一种面向实际操作的测量和评估方法,以实现长期质量控制。该方法能够发现参数的突然变化,并以简单精确的方式确定长期漂移效应,其幅度在测量变化范围内,周期长达数月。在六个月的时间里,不仅可以确定具有长时间周期且幅度在千分之一范围内的漂移效应,还能确定变异系数小于0.3%的情况。