Suppr超能文献

Detection of traces of formaldehyde in pure air by gas chromatography and helium ionization detection.

作者信息

Andrawes F F

出版信息

J Chromatogr Sci. 1984 Nov;22(11):506-8. doi: 10.1093/chromsci/22.11.506.

Abstract

The sensitive helium ionization detector (HID) was used for the direct determination of ppm to ppb levels of formaldehyde in air. Two methods to generate formaldehyde in an air stream were evaluated. The first method utilized a paraformaldehyde permeation tube and the second utilized a motor driven syringe and a dilute solution of formaldehyde. The two methods were evaluated using gas chromatography with HID and spectrophotometry. The paraformaldehyde permeation tube generates only about 60% of the theoretical value, while the motor driven syringe was accurate for levels below 2 ppm; however, at a concentration of 13 ppm or above, oligomers or other chemical forms of CH2O are formed to decrease the concentration of gaseous CH2O produced.

摘要

相似文献

1
Detection of traces of formaldehyde in pure air by gas chromatography and helium ionization detection.
J Chromatogr Sci. 1984 Nov;22(11):506-8. doi: 10.1093/chromsci/22.11.506.
5
Determination of benzoic acid in air.
Am Ind Hyg Assoc J. 1984 Oct;45(10):727-30. doi: 10.1080/15298668491400502.
6
[Gas chromatography for determination of N-isopropylaniline in workplace atmosphere].
Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi. 2015 Feb;33(2):143-5.