Suppr超能文献

[Gas chromatography for determination of N-isopropylaniline in workplace atmosphere].

作者信息

Xuan Jing, Pan Xingfu, Liu Xiaodong, Yan Huifang, Zhang Yanshu

机构信息

School of Public Health, Hebei United University, Tangshan 063000, China.

E-mail:

出版信息

Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi. 2015 Feb;33(2):143-5.

Abstract

OBJECTIVE

To establish a method for determination of N-isopropylaniline in the workplace atmosphere by gas chromatography.

METHODS

Air samples were collected by silica gel tube and desorbed by acetone. Then they were separated through DB-WAX columns and N-isopropylaniline was determined by flame ionization detector.

RESULTS

The concentration of N-isopropylaniline showed a good linear relationship within the range of 1.40∼665.0 µg/ml. The sampling efficiency was 100%. The accuracy was 96%∼ 99% and the precision was 2.1%∼7.0%. The minimum detectable concentration was 0.056 mg/m(3) (with sampled air volume of 7.5 L).

CONCLUSION

The method meets the requirements of analysis and applies to the determination of N-isopropylaniline in the workplace atmosphere.