Suppr超能文献

使用三氟乙酸酐,采用填充柱和毛细管柱气相色谱法及气相色谱/质谱联用法对苯丙胺快速进样口衍生化进行定量研究。

Quantitative investigation of rapid injector port derivatization of amphetamine using trifluoroacetic anhydride with packed and capillary column GC and GC/MS methods.

作者信息

Eiceman G A, Leasure C S, Selim S L

出版信息

J Chromatogr Sci. 1984 Nov;22(11):509-13. doi: 10.1093/chromsci/22.11.509.

Abstract

Direct injector port derivatization of amphetamine through coinjection of sample with trifluoroacetic anhydride was quantitatively investigated and found to be a rapid and more sensitive alternative to longer solution-based derivatization procedures. Under optimum conditions determined in this work, limits of detection were 1 ng using gas chromatography with flame ionization detection and 1 pg using single ion monitoring of ion 140.0 amu (M-91+) in gas chromatography/mass spectrometry methods. The previously reported derivatization period of over 12 hr was greatly reduced using instantaneous derivatization with a 100- to 1000-fold increase in limits of detection. Effects of inlet temperature, mass of trifluoroacetic anhydride, and column pre-treatment on chromatographic performance are reported. Application to analysis of serum from amphetamine-dosed rats was used for comparison to accepted solution-based derivatization.

摘要

通过将样品与三氟乙酸酐共注入对苯丙胺进行直接进样口衍生化,并进行了定量研究,结果发现这是一种快速且比基于溶液的更长衍生化程序更灵敏的替代方法。在本研究确定的最佳条件下,使用带火焰离子化检测的气相色谱法时检测限为1 ng,使用气相色谱/质谱法对离子140.0 amu(M-91+)进行单离子监测时检测限为1 pg。使用瞬时衍生化大大缩短了之前报道的超过12小时的衍生化时间,检测限提高了100至1000倍。报告了进样口温度、三氟乙酸酐质量和柱预处理对色谱性能的影响。将其应用于分析给予苯丙胺的大鼠血清,并与公认的基于溶液的衍生化方法进行比较。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验