Compton scatter with polychromatic sources for lung densitometry.

作者信息

Hanson J A, Moore W E, Figley M M, Duke P R

出版信息

Med Phys. 1984 Sep-Oct;11(5):633-7. doi: 10.1118/1.595558.

Abstract

A mobile lung densitometer using conventional x-ray tubes, NaI detectors, and principles of two-source, two-detector Compton scattered densitometry, is described. The device is capable of one to two per second density measurements from a 45-cm3 volume with a precision of 5%. The expected in vivo accuracy (2%-3%) is determined by using an anthropomorphic phantom with replaceable lung inserts. The unintentional detection of multiple-scattered x rays results in a small density-dependent error. This error is predictable and relatively insensitive to differences in surrounding absorbers such as the chest wall. With this device, dynamic in vivo densitometry of the lung in the clinical laboratory and intensive care unit will be possible.