Energy imparted to the patient in diagnostic radiology: calculation of conversion factors for determining the energy imparted from measurements of the air collision kerma integrated over beam area.

作者信息

Alm Carlsson G, Carlsson C A, Persliden J

出版信息

Phys Med Biol. 1984 Nov;29(11):1329-41. doi: 10.1088/0031-9155/29/11/002.

Abstract

The energy imparted to the patient in diagnostic radiology, related to radiation risk in examinations of the trunk and head, can be deduced from a measurement of the air collision kerma (or exposure) of the incident primary photons integrated over beam area by using a thin, flat ionisation chamber covering the entire roentgen beam. Factors for converting the integral of the air collision kerma to energy imparted to the patient have been calculated using a Monte Carlo method. The patient is simulated by laterally infinite water slabs with thicknesses from 100-300 mm. Calculations are performed for monoenergetic photons (5-300 keV) and energy spectra commonly used in diagnostic radiology (40-130 kV acceleration potential differences and values of the half-value thickness of air collision kerma in aluminium from 0.9 to 9.9 mm). Correction factors which take into account the additional escape of scattered photons from the sides of a laterally finite water slab as a function of field size and focal distance are also given.