Gramiak R, Holen J
Ultrasound Med Biol. 1984 Mar-Apr;10(2):215-24. doi: 10.1016/0301-5629(84)90220-5.
An independent directional Doppler system operating in pulsed or continuous-wave mode and a real-time spectral analyzer with a 3 ms temporal resolution have been used to supplement conventional real-time and M-mode echocardiography. Examination techniques as well as findings in normal and abnormal hearts are presented. Continuous-wave Doppler is most useful for detection of peak frequency shifts for calculation of valvular gradients while pulsed Doppler is superior in the detection of small jets of abnormal flow and in display of spectral broadening. Real-time spectral analysis allows interactive selection of peak frequency shifts, can separate coincident flows by signal intensity and depicts high frequency valvular movements to rival the capabilities of M-mode.
一个以脉冲或连续波模式运行的独立定向多普勒系统和一个具有3毫秒时间分辨率的实时频谱分析仪已被用于补充传统的实时和M型超声心动图。本文介绍了检查技术以及正常和异常心脏的检查结果。连续波多普勒在检测用于计算瓣膜压力阶差的峰值频移方面最为有用,而脉冲多普勒在检测异常血流的小喷射流和显示频谱增宽方面更具优势。实时频谱分析允许交互式选择峰值频移,可通过信号强度分离重叠的血流,并描绘高频瓣膜运动,其功能可与M型超声心动图相媲美。