Yamanashi W S, Wheatley K K, Lester P D, Anderson D W
Physiol Chem Phys Med NMR. 1984;16(3):237-50.
Various artifacts of Magnetic Resonance Imaging (MRI) typically associated with currently available imaging techniques such as projection reconstruction and two-dimensional fourier transform (2D-FT) are described and illustrated. Examples of MRI artifacts were obtained with an imaging unit with a super conducting magnet operated at .15 Tesla and .27 Tesla with corresponding proton resonance frequency of 6.4 MHz and 11.25 MHz. The .15 Tesla images were obtained using projection reconstruction and the .27 Tesla using the 2D-FT method. Instrument related artifacts include those due to direct current (DC), projection, gradient offset, active shimming, phase encoding, and pulse sequencing. Other often encountered artifacts are related to the patient. These include those due to motion, ferromagnetic effect, and tissue contents. The cause of these artifacts and how (if possible) they may be eliminated or minimized is discussed.
描述并举例说明了各种通常与当前可用成像技术(如投影重建和二维傅里叶变换(2D - FT))相关的磁共振成像(MRI)伪影。MRI伪影的示例是通过一台超导磁体的成像设备获得的,该磁体分别在0.15特斯拉和0.27特斯拉下运行,相应的质子共振频率为6.4兆赫兹和11.25兆赫兹。0.15特斯拉的图像是使用投影重建获得的,0.27特斯拉的图像是使用2D - FT方法获得的。与仪器相关的伪影包括由直流(DC)、投影、梯度偏移、主动匀场、相位编码和脉冲序列引起的伪影。其他经常遇到的伪影与患者有关。这些包括由运动、铁磁效应和组织成分引起的伪影。讨论了这些伪影的成因以及如何(如果可能的话)消除或最小化它们。