[Radiation dose in renal magnification angiography compared with conventional angiography (author's transl)].

作者信息

Maly P, Knöös T

出版信息

Radiologe. 1978 May;18(5):175-8.

PMID:652984
Abstract

Measurements of skin dose, integrated skin dose area, and integral dose in combined conventional and magnification renal angiography using lanthanoxibromide screens MR 600 and carbon fibre plate in filmchangertop have been made. The mean skin dose in magnification technique was 0.56 rad./film with focus-skin distance of 35 cm, FFD 1 m, 85-90 kV and sagittal diameter of abdomen of 20 cm. The combination of angiographic technique with and without magnification and the use of a high-sensitivity receiving unit makes it possible to keep the total skin dose per investigation (8.5 rad.) within the limits accepted in conventional abdominal angiography.

摘要

已对在使用溴氧化镧屏幕MR 600和胶片更换器顶部的碳纤维板进行的传统与放大肾血管造影联合检查中,皮肤剂量、皮肤剂量积分面积和积分剂量进行了测量。在放大技术中,当焦皮距为35厘米、源皮距为1米、管电压85 - 90千伏且腹部矢状径为20厘米时,平均皮肤剂量为0.56拉德/片。血管造影技术的放大与非放大相结合以及使用高灵敏度接收单元,使得每次检查的总皮肤剂量(8.5拉德)能够保持在传统腹部血管造影可接受的限度内。