Bojsen J, Staberg B, Kølendorf K
Clin Physiol. 1984 Aug;4(4):309-20. doi: 10.1111/j.1475-097x.1984.tb00806.x.
Portable, solid state, cadmium telluride [CdTe(C1)] detectors were used for measurements of 133Xe-disappearance rate constants as a measure of subcutaneous (s.c.) blood flow. To eliminate the disturbance on the measurements from combined local convection and diffusion of xenon in the s.c. tissue, two methods were demonstrated to be feasible in both control persons and anaesthetized rabbits. A small volume (50-80 microliters) was injected into the s.c. tissue at a depth of 5 mm and the CdTe(C1) detector elevated 20 mm above the skin surface. In this situation identical disappearance rate constants were recorded by the CdTe(C1) detector and the NaI(T1) reference detector at 15-20 cm from the depot. Similiar results were obtained when either a large volume (greater than or equal to 800 microliters of Xe in humans) was injected into the s.c. tissue or the labelling was performed with a smaller volume in an atraumatic uniform manner to obtain a constant Xe concentration exceeding the field of view of the CdTe(C1) detector, which in these cases was attached directly to the skin surface. The coefficients of variation (CV) of the disappearance rate constants measured by the CdTe(C1) detector were 9% and 11%, and these values did not differ significantly from each other (P greater than 0.2), or from CV-values calculated from measurements with NaI(T1) detectors.
便携式固态碲化镉[CdTe(C1)]探测器用于测量133Xe消失速率常数,以此作为皮下血流的指标。为消除皮下组织中氙的局部对流和扩散对测量的干扰,在对照人群和麻醉兔中证明了两种方法可行。将小体积(50 - 80微升)注入到5毫米深度的皮下组织中,CdTe(C1)探测器置于皮肤表面上方20毫米处。在此情况下,CdTe(C1)探测器和距离注射部位15 - 20厘米处的NaI(T1)参考探测器记录到相同的消失速率常数。当将大体积(人类中大于或等于800微升氙)注入皮下组织,或以无创均匀方式用较小体积进行标记以获得超过CdTe(C1)探测器视场的恒定氙浓度(在这些情况下,探测器直接附着在皮肤表面)时,也获得了类似结果。CdTe(C1)探测器测量的消失速率常数的变异系数(CV)分别为9%和11%,这些值彼此之间差异不显著(P大于0.2),与用NaI(T1)探测器测量计算得到的CV值也无显著差异。