Bongard O, Krahenbuhl B
Clin Physiol. 1984 Oct;4(5):393-401. doi: 10.1111/j.1475-097x.1984.tb00124.x.
Pedal transcutaneous partial pressure of oxygen (tcPO2) has been measured by polarographic method, heating the skin at 44 degrees C. In 50 normal subjects, mean tcPO2 measured 54.5 +/- 7 mmHg. Among 43 patients suffering from chronic ischaemia of lower limbs, mean tcPO2 measured 40.8 +/- 8 mmHg in patients with from claudication and 16.1 +/- 15 mmHg in patients suffering from rest pain and/or gangrene. The variability of repeated tcPO2 measurements, expressed as 1 SD of the mean, was 4.5 mmHg in normal subjects and 2.9 mmHg in patients. The relationships between pedal subcutaneous blood flow measured in xenon-133 clearance method and pedal tcPO2 have been studied in nine normal subjects and in five patients suffering from severe chronic ischaemia of lower limbs (rest pain and/or gangrene). There was a positive correlation between blood flow and tcPO2 in normal subjects (r = 0.77, P less than 0.001). In patients suffering from severe ischaemia, there was no correlation between these two parameters, but measured blood flow was sometimes very high in areas where tcPO2 was low. It is likely that 133Xe clearance method considerably overestimates local blood flow in these patients, because there is considerably less fat in subcutaneous tissue of chronic severely ischaemic areas. Thus, partition coefficient should be determined in each patient. However, tcPO2 may constitute an index of nutritional circulation, while 133Xe clearance actually measures total subcutaneous blood flow.
采用极谱法在44摄氏度加热皮肤的情况下测量了经皮氧分压(tcPO2)。在50名正常受试者中,平均tcPO2为54.5±7 mmHg。在43例下肢慢性缺血患者中,间歇性跛行患者的平均tcPO2为40.8±8 mmHg,静息痛和/或坏疽患者的平均tcPO2为16.1±15 mmHg。重复测量tcPO2的变异度(以均值的1个标准差表示)在正常受试者中为4.5 mmHg,在患者中为2.9 mmHg。在9名正常受试者和5例严重下肢慢性缺血(静息痛和/或坏疽)患者中研究了用氙-133清除法测量的足部皮下血流与足部tcPO2之间的关系。正常受试者中血流与tcPO2呈正相关(r = 0.77,P<0.001)。在严重缺血患者中,这两个参数之间无相关性,但在tcPO2较低的区域有时测量到的血流非常高。很可能氙-133清除法在这些患者中大大高估了局部血流,因为慢性严重缺血区域的皮下组织中脂肪明显减少。因此,应在每个患者中确定分配系数。然而,tcPO2可能构成营养循环的一个指标,而氙-133清除法实际上测量的是皮下总血流。