Venous air embolism through infusion sets. Theoretical considerations, model experiments and prevention.

作者信息

Pedersen N T, Hessov I

出版信息

Acta Anaesthesiol Scand. 1978;22(2):117-22. doi: 10.1111/j.1399-6576.1978.tb01288.x.

DOI:10.1111/j.1399-6576.1978.tb01288.x
PMID:654852
Abstract

Some infusion sets are so designed that, even when infusing into peripheral veins, a negative pressure may occur in a large portion of the infusion tubing, with the result that air enters the system from a side drip which has run dry, or though leaks. In model experiments, it was found that, when infusing into a peripheral vein, volumes of air up to 13 ml per min can enter the system from an empty side drip. From an analysis of the pressure conditions in infusion sets before and during infusion, it is concluded that infusions sets should be so designed that clamps and connections forthe attachment of a side drip are placed distally on the infusion tubing. Even then there is a risk of air embolism. The tubes for both the main and the side drip should therefore be so long that each of them can be fixed below the heart level of the patient, thus forming a waterlock and preventing the entry of air.

摘要

相似文献

1
Venous air embolism through infusion sets. Theoretical considerations, model experiments and prevention.
Acta Anaesthesiol Scand. 1978;22(2):117-22. doi: 10.1111/j.1399-6576.1978.tb01288.x.
2
A universal air-bubble detector for infusion pumps.
J Med Eng Technol. 1984 Mar-Apr;8(2):64-5. doi: 10.3109/03091908409036745.
3
[Nil nocere. Possibility of air embolism during i.v. infusion].
Z Arztl Fortbild (Jena). 1971 Jun 15;65(12):588.
4
Needle venting of air from intravenous tubing.通过针头排出静脉输液管中的空气。
Anesth Analg. 1974 Nov-Dec;53(6):1016-7. doi: 10.1213/00000539-197453060-00040.
5
[Micropore filters for preventing venous air embolism during infusions (author's transl)].
Anasth Intensivther Notfallmed. 1982 Feb;17(1):58-61.
6
[Causes of gas embolism with the use of apparatus for drip infusion].[使用滴注输液装置时气体栓塞的原因]
Sven Lakartidn. 1956 Oct 26;53(43):2803-13.
7
[Possible solutions to the problems of pressure transfusion (infusion) from bottles].[瓶装压力输血(输液)问题的可能解决方案]
Infusionstherapie (1973). 1973 Oct;1(1):58-61.
8
[Standards of infusion pumps].
Klin Anasthesiol Intensivther. 1977(14):139-51.
9
Venous air embolism during home infusion therapy.家庭输液治疗期间的静脉空气栓塞
Pediatrics. 2002 Jan;109(1):E15. doi: 10.1542/peds.109.1.e15.
10
Massive air emboli from intravenous infusion pump: etiology and prevention.
Am J Surg. 1979 Feb;137(2):274-5. doi: 10.1016/0002-9610(79)90163-6.

引用本文的文献

1
Direct blood-pressure measurements: risks, technological evolution and some current problems.
Med Biol Eng Comput. 1984 Jan;22(1):2-5. doi: 10.1007/BF02443737.