Mutschler W, Claes L, Mohr W, Hutzschenreuter P
Arch Orthop Trauma Surg (1978). 1978 Apr 18;91(2):129-35. doi: 10.1007/BF00378895.
The screws and plates used in operative treatment of fractures cause moments of friction between the bone screw head and the bone plate hole during the tightening of the screws. This results in hidden damage to the grinding metal surfaces, with the danger of erosion corrosion and in a reduction in the screw force. By covering the screw heads with the lubricating sprays PTFE or graphite friction was significantly reduced, the screw force increased by 40% and extensive damage to the surfaces avoided. Lubricating sprays or their extracts, in fetal calf serum or in dimethylsulfoxide, were added to fibroblast cultures. Toxicity was manifest as changes in the synthesis of DNA, as measured by autoradiography. Sprays or spray extracts in fetal calf serum did not inhibit cell proliferation. In vivo experiments in rats reflected the in vitro results, there being no evidence that toxic substances were released from the sprays by physiological solvents: There was a lack of histopathological changes in the inner organs and a local encapsulation of implants which was essentially like that in a teflon control group.
骨折手术治疗中使用的螺钉和钢板在拧紧螺钉时会导致骨螺钉头部与骨板孔之间产生摩擦力矩。这会对磨削金属表面造成隐性损伤,存在腐蚀侵蚀的风险,并会降低螺钉的力。通过用聚四氟乙烯(PTFE)或石墨润滑喷雾覆盖螺钉头部,摩擦力显著降低,螺钉力增加了40%,并避免了表面的广泛损伤。将润滑喷雾或其提取物添加到胎牛血清或二甲基亚砜中的成纤维细胞培养物中。通过放射自显影测量,毒性表现为DNA合成的变化。胎牛血清中的喷雾或喷雾提取物不会抑制细胞增殖。在大鼠身上进行的体内实验反映了体外实验结果,没有证据表明生理溶剂会从喷雾中释放出有毒物质:内脏没有组织病理学变化,植入物的局部包封基本上与聚四氟乙烯对照组相似。