Twenty-three episodes of pseudobacteremia occurring over a seven-month period were traced to cross-contamination by the automated blood culture analyzer (BACTEC 460) used in the microbiology laboratory. An epidemiologic investigation of an unusual cluster of three patients with pseudobacteremia caused by oxacillin-resistant Staphylococcus aureus led to the identification of the problem. Mock trials of the blood culture procedure confirmed that the blood culture analyzer was the source of contamination. After the needle sterilizer was replaced by the manufacturer, the problem of cross-contamination abated. Contamination of sterile blood cultures by an instrument intended to identify bacteremia rapidly may lead to incorrect diagnosis, unnecessary administration of antibiotics and prolonged hospitalization. Because of the widespread use of automated blood culture analyzers in the US, physicians, microbiologists, and infection control personnel should be alert to the possibility of cross-contamination and the subtle way in which it may present.

摘要

在七个月的时间里发生的23起假菌血症事件被追溯到微生物实验室使用的自动血培养分析仪(BACTEC 460)的交叉污染。对由耐苯唑西林金黄色葡萄球菌引起的三例假菌血症患者的异常聚集进行的流行病学调查导致了问题的发现。血培养程序的模拟试验证实血培养分析仪是污染源。在制造商更换针头消毒器后,交叉污染问题得到缓解。旨在快速识别菌血症的仪器对无菌血培养的污染可能导致错误诊断、不必要的抗生素使用和住院时间延长。由于自动血培养分析仪在美国的广泛使用,医生、微生物学家和感染控制人员应警惕交叉污染的可能性及其可能呈现的微妙方式。