Vannier A M, Tarrand J J, Murray P R
Division of Laboratory Medicine, Washington University School of Medicine, St. Louis, Missouri 63110.
J Clin Microbiol. 1988 Sep;26(9):1867-8. doi: 10.1128/jcm.26.9.1867-1868.1988.
Pseudobacteremias in blood cultures performed on the BACTEC radiometric blood culture system have been reported. We report three cases of cross contamination with Mycobacterium avium that occurred when mycobacteria were cultured with the BACTEC 460 TB system. Malfunction of the needle sterilization heating block was demonstrated.