使用MS-2系统快速检测耐甲氧西林葡萄球菌。

Rapid detection of methicillin-resistant staphylococci with the MS-2 system.

作者信息

Flournoy D J, Hussain Qadri S M, Johnson W S

出版信息

Can J Microbiol. 1984 Apr;30(4):488-90. doi: 10.1139/m84-071.

Abstract

The ability of the updated Abbott MS-2 system to detect methicillin-resistant Staphylococcus aureus (MRSA) was compared with disk agar diffusion and broth microdilution. Of the 87 MRSA isolates tested, the MS-2 system correctly detected 85 (97.7%) in 5 h. Seventy-two of the isolates were detected by disk agar diffusion within 24 h, and 15 more by 48 h. Broth microdilution detected 71 MRSA by 24 h of incubation and 11 more by 48 h. The updated MS-2 system yielded reliable results that were highly comparable with the standard techniques.

摘要

将更新后的雅培MS-2系统检测耐甲氧西林金黄色葡萄球菌(MRSA)的能力与纸片琼脂扩散法和肉汤微量稀释法进行了比较。在检测的87株MRSA分离株中,MS-2系统在5小时内正确检测出85株(97.7%)。72株分离株在24小时内通过纸片琼脂扩散法检测出,另有15株在48小时内检测出。肉汤微量稀释法在培养24小时时检测出71株MRSA,在48小时时又检测出11株。更新后的MS-2系统产生了可靠的结果,与标准技术高度可比。