Henry D, Kunzer L, Ngui-Yen J, Smith J
J Clin Microbiol. 1986 Apr;23(4):718-24. doi: 10.1128/jcm.23.4.718-724.1986.
A study was undertaken to compare four commercial systems for testing the antimicrobial susceptibility patterns of gram-positive cocci. The reference method was an agar dilution method. The systems evaluated were the MS-2 system (Abbott Diagnostics Div., Mississauga, Ontario), the AutoMicrobic system (AMS) (Vitek Systems, Inc., Hazelwood, Mo.) with the gram-positive susceptibility (GPS) card, the Sceptor system (BBL Microbiology Systems, distributed by Becton Dickenson, Canada Inc., Mississauga, Ontario), and the Micro-Media system (Beckman Instruments, Inc., Anaheim, Calif.). There was a greater than 98% essential accord (EA) between all test results and the reference method results when testing 134 isolates of Staphylococcus aureus. In testing 79 isolates of coagulase-negative staphylococci the EA was greater than 97% with all systems except the MS-2. In the MS-2 system only, 30% of tests were interrupted by the instrument because of insufficient growth in the control chamber. Excluding the Sceptor system, the EA was greater than 96% on testing 70 isolates of enterococcus. In testing 15 isolates of group B Streptococcus there was 91% EA with the AMS and Sceptor systems and only 71 and 88% EA with the MS-2 and Micro-Media systems, respectively. The new AMS GPS MIC card was tested against 29 methicillin-resistant S. aureus, 10 coagulase-negative staphylococci, and 9 enterococci, and it gave more accurate results than the earlier GPS breakpoint card. The Micro-Media and MS-2 systems did not reliably detect marginally methicillin-resistant S. aureus. The MS-2 was the least expensive system to operate on a cost per test basis ($3.59 Can.), whereas the Sceptor was the most expensive system ($5.29 Can.). The AMS ws the least labor intensive (0.9 min per test), and the Sceptor system was the most time consuming (2.9 min per test).
开展了一项研究,以比较用于检测革兰氏阳性球菌抗菌药敏模式的四种商用系统。参考方法为琼脂稀释法。评估的系统包括MS-2系统(雅培诊断部,安大略省密西沙加市)、配备革兰氏阳性药敏(GPS)卡的自动微生物系统(AMS)(维泰克系统公司,密苏里州黑兹尔伍德市)、Sceptor系统(BBL微生物系统公司,由加拿大贝克顿·迪金森公司分销,安大略省密西沙加市)以及微量培养基系统(贝克曼仪器公司,加利福尼亚州阿纳海姆市)。在检测134株金黄色葡萄球菌时,所有检测结果与参考方法结果之间的基本符合率(EA)大于98%。在检测79株凝固酶阴性葡萄球菌时,除MS-2系统外,所有系统的EA均大于97%。仅在MS-2系统中,30%的检测因对照室生长不足而被仪器中断。排除Sceptor系统,在检测70株肠球菌时,EA大于96%。在检测15株B族链球菌时,AMS和Sceptor系统的EA为91%,而MS-2和微量培养基系统的EA分别仅为71%和88%。新型AMS GPS MIC卡针对29株耐甲氧西林金黄色葡萄球菌、10株凝固酶阴性葡萄球菌和9株肠球菌进行了测试,其结果比早期的GPS断点卡更准确。微量培养基系统和MS-2系统不能可靠地检测出轻度耐甲氧西林金黄色葡萄球菌。以每次检测成本计算,MS-2系统是操作成本最低的系统(3.59加元),而Sceptor系统是最昂贵的系统(5.29加元)。AMS的劳动强度最小(每次检测0.9分钟),而Sceptor系统耗时最长(每次检测2.9分钟)。