Suppr超能文献

Effect of protein on the determination of total bile acids in serum.

作者信息

Hanson N Q, Freier E F

出版信息

Clin Chem. 1983 Jan;29(1):171-5.

PMID:6571720
Abstract

We measured total serum bile acids on a fluorescence-light-scattering micro centrifugal analyzer by the direct enzymatic method with 3 alpha-hydroxysteroid dehydrogenase (EC 1.1.1.50) and with resazurin as a fluorogenic electron acceptor. We found that serum protein has an inhibitory effect on the measurement of bile acids, but this effect was eliminated by adding bovine serum albumin to the reaction mixture in a final protein concentration (12.2 g/L) that was high compared with that contributed by a normal serum specimen. The assay is a sensitive method that reaches equilibrium in 5 min. The method is microscale (5 microL of sample, 150 microL of working reagent), is easy to perform, and is accurate (analytical recovery = 104.1%) and precise (CV = 11.1 and 5.7% on specimens with bile acid concentrations of 7.6 and 35.4 mumol/L, respectively). Normal values are 1-12 and less than 9 mumol/L on nonfasting and fasting individuals, respectively. Pure 3 alpha-hydroxysteroid dehydrogenase must be used: we found several enzyme preparations that gave falsely high values for bile acid.