Kubiatowicz D O, Ithakissios D S, Wicks J H, Heerwald P E
J Nucl Med. 1978 Jul;19(7):854-7.
We describe a T3 uptake test that uses magnetic albumin microparticles to separate free from bound [125I] T3. The test uses 0.1 ml serum in a total assay volume of 1.1 ml, and is performed in barbital buffer of pH 7.6, ionic strength 0.2. The accuracy of the test is not affected by the following ranges: volume, 1-3 ml; incubation time, 15-120 min; and incubation temperature, 4-35 degrees C. We describe a magnetic separator device that allows simultaneous handling of 100 assay tubes with minimum manipulation. The intra-assay and interassay coefficients of variance were 2.87% and 5.20%, respectively.
我们描述了一种使用磁性白蛋白微粒从结合的[125I]T3中分离游离T3的T3摄取试验。该试验在1.1ml的总测定体积中使用0.1ml血清,在pH7.6、离子强度0.2的巴比妥缓冲液中进行。试验的准确性不受以下范围影响:体积1 - 3ml;孵育时间15 - 120分钟;孵育温度4 - 35摄氏度。我们描述了一种磁性分离装置,该装置允许以最少的操作同时处理100个测定管。批内和批间变异系数分别为2.87%和5.20%。