Hart H R, Bottomley P A, Edelstein W A, Karr S G, Leue W M, Mueller O, Redington R W, Schenck J F, Smith L S, Vatis D
AJR Am J Roentgenol. 1983 Dec;141(6):1195-201. doi: 10.2214/ajr.141.6.1195.
The choice of the strength of the magnetic field for an imaging system based on the nuclear magnetic resonance of hydrogen is considered. It is shown by an analysis based on in vitro data that the quality, or contrast-to-noise ratio, of images based on T1 or T2 discrimination increases with field up to 1.5-2 T. After a brief discussion of potential high-field limitations, results are presented which show that images of the human head with excellent anatomic detail can be produced at 1.5 T or 64 MHz.
考虑了基于氢核磁共振的成像系统磁场强度的选择。基于体外数据的分析表明,基于T1或T2分辨的图像质量(即对比度噪声比)随着磁场强度增加到1.5 - 2T而提高。在简要讨论了潜在的高场限制之后,给出的结果表明,在1.5T或64MHz下可以生成具有出色解剖细节的人头图像。