Suppr超能文献

达德自动纤维蛋白原测定仪(Dade Auto-FI)和生物探索纤维蛋白仪(Bio-Quest Fibrometer)上凝血酶原时间和活化部分凝血活酶时间的比较

Comparison of prothrombin time and activated partial thromboplastin time on Dade Auto-FI and Bio-Quest Fibrometer.

作者信息

Villa L, Ibanez J, Castro A

出版信息

Res Commun Chem Pathol Pharmacol. 1978 Apr;20(1):165-73.

PMID:663397
Abstract

Plasma assay results for prothrombin time and activated partial thromboplastin time obtained on the fully automated Dade Auto-FI were compared with those obtained on the more traditional semi-automated Bio-Quest Fibrometer. The precision data showed in a majority of cases the superior reproducibility of the Auto-FI over the fibrometer. However, we found that results of APTT utilizing both methodologies were not at present interchangeable.

摘要

将在全自动达德自动凝血仪(Dade Auto-FI)上获得的凝血酶原时间和活化部分凝血活酶时间的血浆检测结果,与在更为传统的半自动生物探索纤维蛋白仪(Bio-Quest Fibrometer)上获得的结果进行了比较。精密度数据表明,在大多数情况下,自动凝血仪(Auto-FI)的重复性优于纤维蛋白仪。然而,我们发现目前使用这两种方法获得的活化部分凝血活酶时间(APTT)结果不可互换。