Suppr超能文献

Characterization of tissue via coherent-to-Compton scattering ratio: sensitivity considerations.

作者信息

Karellas A, Leichter I, Craven J D, Greenfield M A

出版信息

Med Phys. 1983 Sep-Oct;10(5):605-9. doi: 10.1118/1.595330.

Abstract

It is known that the ratio (R) of the detected coherent and Compton scattered photons from bone can be used in order to determine its mineral density. This technique utilizes the dependence of the coherent scattering on the effective atomic number (Z) of the scattering medium. It is generally accepted that a small scatter angle is preferred in order to ensure adequate counting statistics by favoring the detection of more coherent photons. Moreover, it has been assumed that a change in the scatter angle does not affect the sensitivity of the measurement. Our theoretical calculations for 60-keV photons and for the range of Z that corresponds to trabecular bone, indicate that increasing the scatter angle results in a stronger power dependence of the measured ratio on Z. This implies that by increasing the scatter angle, smaller changes in the mineral density can be detected, thus improving the sensitivity of the measurement. This effect was investigated experimentally by using a collimated beam of 59.54-keV photons from Am-241 (44.4 GBq) and a collimated intrinsic germanium detector. Solutions of K2HPO4 with different concentrations were used in order to simulate trabecular bone. The scatter spectra were recorded for all solutions at six scatter angles between 37 degrees and 98 degrees and the value of R was computed for each spectrum. The sensitivity of the measurement, evaluated from these experiments increased, with the increase of the scatter angle.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验