A new technique of internal fixation of the flail chest was tested successfully in sheep. After reduction a 0.8 mm wire is passed twice intramedullary across the fracture and twisted under tension with two loops on the external side of the rib. This wire neutralizes bending and rotational forces and produces axial interfragmental compression. Combined with tension band wire fixation of an eventually fractures sternum it produced sufficient stability even in the very thin parasternal part of ribs. No respiratory support was necessary postoperatively. The method is simple and needs no special equipment.
一种连枷胸内固定新技术在绵羊身上成功进行了测试。复位后,将一根0.8毫米的钢丝两次经髓内穿过骨折部位,并在肋骨外侧用两个环在张力下扭转。这根钢丝可抵消弯曲和旋转力,并产生轴向骨折块间加压。结合最终骨折胸骨的张力带钢丝固定,即使在肋骨胸骨旁非常薄的部位也能产生足够的稳定性。术后无需呼吸支持。该方法简单,不需要特殊设备。