Gross W, Festge B, Festge O A, Hajdu K, Rosenbaum K D
Arch Gynecol. 1983;234(1):67-73. doi: 10.1007/BF02114728.
To make manometric studies of urethral function and obtain exact recordings of the so-called stress profile requires a pressure rise rate which exceeds that of the pressure complex due to coughing. The syringe perfusion pump commonly used in urodynamics limits the pressure rise rate. The rate ist also reduced by catheters with high compliance. A suitable recording accuracy necessitates values greater than 150 kPa/s. This can be achieved with the aid of an improved pneumatic-hydraulic infusion system and low-compliance catheters. We describe the improvement of a recording system and demonstrate the importance of a suitable pressure rise rate.
要对尿道功能进行测压研究并获得所谓应力曲线的精确记录,需要压力上升速率超过咳嗽引起的压力波的上升速率。尿动力学中常用的注射器灌注泵限制了压力上升速率。高顺应性导管也会降低该速率。合适的记录精度需要大于150 kPa/s的值。借助改进的气动-液压输注系统和低顺应性导管可以实现这一点。我们描述了一种记录系统的改进,并证明了合适的压力上升速率的重要性。