Feld L G, Langford D J, Schwartz G J
Clin Pediatr (Phila). 1984 Mar;23(3):154-6. doi: 10.1177/000992288402300304.
Bilirubin interferes with the measurement of plasma creatinine by the kinetic colorimetric Jaffe reaction, the method currently used by many hospital laboratories. The purpose of the present study was to evaluate the influence of unconjugated bilirubin on the measurement of plasma creatinine by the AutoAnalyzer Jaffe end point method. This colorimetric method is slower but more precise and accurate in the range of 0.25 to 0.5 mg/dl. We found that bilirubin from jaundiced neonates and from stock solutions did not affect the determination of creatinine chromogen in plasma or in saline, even at concentrations as high as 25 mg/dl. We conclude that the use of the Jaffe end point AutoAnalyzer method with a blank run before each sample will provide an accurate measurement of creatinine in the plasma of the neonate with unconjugated hyperbilirubinemia.