Attiyat A S, Christian G D
Clin Chim Acta. 1984 Feb 28;137(2):151-7. doi: 10.1016/0009-8981(84)90175-x.
Flow injection analysis-atomic absorption (FIA-AA) determination of zinc in human serum samples is demonstrated. The same samples were analyzed for zinc by conventional atomic absorption spectrometry and no significant difference was found between the two methods. Proteins were precipitated from samples by adding equal volumes of 25% (w/v) trichloroacetic acid and 100-microliter aliquots of the protein-free-filtrate were injected. Precision was +/-3.47% relative standard deviation. The frequency of measurements can be as high as 180 per hour. No pump was used, only the negative pressure from the nebulizer was utilized to aspirate the carrier stream. Simple 1:5 dilution of sample with water, followed by FIA-AA measurements, gave higher zinc levels. The effect of sample volume and the length of dispersion coil on the FIA signal was studied.
本文展示了流动注射分析-原子吸收光谱法(FIA-AA)测定人血清样本中的锌。采用常规原子吸收光谱法对相同样本进行锌分析,两种方法之间未发现显著差异。通过加入等体积的25%(w/v)三氯乙酸使样本中的蛋白质沉淀,然后注入100微升无蛋白滤液的等分试样。精密度为相对标准偏差±3.47%。测量频率可达每小时180次。未使用泵,仅利用雾化器的负压抽吸载流。样本用水简单1:5稀释后,再进行FIA-AA测量,得到的锌含量更高。研究了样本体积和分散盘管长度对FIA信号的影响。