Suppr超能文献

全人工心脏置换的实验研究

Experimental studies in total artificial heart replacement.

作者信息

Iwaya F, Hoshino S, Igari T, Honda K

出版信息

Jpn Circ J. 1984 Mar;48(3):312-8. doi: 10.1253/jcj.48.312.

Abstract

Twenty-seven calves were operated on in our institute for total artificial heart replacement. Thomasu hearts (made by the Thomasu Technical Company, Japan) were used in 21 calves from February 1980 to May 1983. Utah Jarvik 7 hearts were used for the first five calves and the 17th calf. The air-driven Thomasu heart, made of segmented polyurethane, has 95 ml and 85 ml of stroke volume in its left and right sides respectively and showed recognizable function curves in mock circulation systems. There were no satisfactory results in the first 19 cases but the 20th calf survived for 28 days. In 1982, we adopted a new surgical method in which, instead of suturing, a sutureless artificial graft with a stainless steel ring was inserted into the aorta and pulmonary truncus. Since then two out of six calves have survived for more than one month; one of these survived for 66 days. The causes of death after the Thomasu heart replacements were varied. Ten (48%) out of 21 cases had heart trouble and four calves (19%) died through thrombus formation at the inflow and outflow valves. Three calves (14%) died because of mechanical failure of the driving system, two (10%) due to surgical failure, one because of pneumonia, and one due to an abnormal reaction after donor blood transfusion. Our longest surviving calf died after 66 days, because of a broken left blood diaphram though, it remained in excellent condition until its death. There was no thrombus formation in either ventricle, no fibrin net formation and no calcium deposit on the surface of the diaphragm.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

在我们研究所,27头小牛接受了全人工心脏置换手术。1980年2月至1983年5月,21头小牛使用了托马苏心脏(由日本托马苏技术公司制造)。前五头小牛和第17头小牛使用了犹他·贾维克7型心脏。气动的托马苏心脏由分段聚氨酯制成,左右心室的搏出量分别为95毫升和85毫升,在模拟循环系统中显示出可识别的功能曲线。前19例手术效果不佳,但第20头小牛存活了28天。1982年,我们采用了一种新的手术方法,即不进行缝合,而是将带有不锈钢环的无缝合人工移植物插入主动脉和肺动脉干。从那时起,六头小牛中有两头存活了一个多月;其中一头存活了66天。托马苏心脏置换术后的死亡原因各不相同。21例中有10例(48%)出现心脏问题,4头小牛(19%)因流入和流出瓣膜处形成血栓而死亡。3头小牛(14%)死于驱动系统的机械故障,2头(10%)死于手术失败,1头死于肺炎,1头死于输入供血后的异常反应。我们存活时间最长的小牛在66天后死亡,原因是左血隔膜破裂,不过,直到死亡前它的状况一直良好。两个心室均未形成血栓,未形成纤维蛋白网,隔膜表面也未出现钙沉积。(摘要截取自250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验