Rocks B F, Sherwood R A, Riley C
Ann Clin Biochem. 1984 Jan;21 ( Pt 1):51-6. doi: 10.1177/000456328402100109.
The use of flow-injection analysis for the direct determination of calcium and magnesium in blood serum and plasma is described. An inexpensive rotary valve is used to inject the serum sample (4 microliters) into a flowing non-segmented stream of reagent which carries the sample slug through a long narrow-bore coil--where it gradually disperses--and into the nebuliser of an atomic absorption spectrometer. This on-stream sample dilution removes the need for predilution of the sample. The resulting absorbance signals are recorded as peaks less than 40 seconds after sample injection. Analytical recoveries and precision are good for both elements and the results by flow-injection analysis compare well with established routine methods.
本文描述了使用流动注射分析法直接测定血清和血浆中的钙和镁。使用一种廉价的旋转阀将血清样品(4微升)注入流动的非分段试剂流中,该试剂流将样品段输送通过一个长的窄孔盘管(样品在其中逐渐分散),然后进入原子吸收光谱仪的雾化器。这种在线样品稀释消除了对样品进行预稀释的需要。在样品注入后不到40秒,所产生的吸光度信号被记录为峰。两种元素的分析回收率和精密度都很好,流动注射分析法的结果与既定的常规方法相比具有良好的可比性。