Suppr超能文献

Validation of Fourier amplitude ratio to quantitate valvular regurgitation.

作者信息

Makler P T, McCarthy D M, Kleaveland J P, Doherty J U, Velchik M G

出版信息

J Am Coll Cardiol. 1984 Jun;3(6):1482-7. doi: 10.1016/s0735-1097(84)80287-9.

Abstract

It has previously been shown that patients with valvular regurgitation can be identified by the ratio of left and right ventricular amplitude values obtained from first harmonic Fourier analysis of the gated blood pool scan. The present study was designed to validate the accuracy of this technique for quantifying the amount of valvular regurgitation. In a blinded analysis of 19 patients who underwent cardiac catheterization, there was a close correlation between the radionuclide and hemodynamic determination of the amount of regurgitation (r = 0.84). The interobserver agreement for calculating the radionuclide data was also high (r = 0.88). These results suggest that the Fourier ventricular amplitude ratio is an accurate and reproducible technique for quantifying valvular regurgitation by gated blood pool scanning.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验