An improvement in the range resolution of ultrasonic pulse echo systems by deconvolution.

作者信息

Carpenter R N, Stepanishen P R

出版信息

J Acoust Soc Am. 1984 Apr;75(4):1084-91. doi: 10.1121/1.390782.

Abstract

Ultrasonic pulse echo systems are often limited in range resolution by the bandwidth of the piezoelectric transducer. Significant improvements in the range resolution of such systems can be obtained by minimizing the effects of the transducer's dynamic response on the overall pulse echo process. An approach to minimize the effects of the transducer is developed from linear system and impulse response techniques. In essence, the pulse echo voltage of interest is deconvolved with a pulse echo reference voltage which is obtained from an air/water interface in the nearfield of the transducer . A computer study of the pulse echo process and the deconvolution process is presented to illustrate the nature of the improvement in range resolution for several cases of interest. Finally, experimental results are presented to illustrate the improvement using commercially available transducers.