Lohman J J, Merkus F W, Costongs G M, Hegtermans E P, Hooymans P M
Pharm Weekbl Sci. 1984 Apr 27;6(2):91-5. doi: 10.1007/BF01953961.
The usefulness of three simple, commercially available ultrafiltration systems [ Centriflo CF25 (A), Centriflo CF50A (B) and Ultrafree (C)] in determining concentrations of unbound carbamazepine in plasma was tested. The concentrations of carbamazepine measured in the first fractions of the ultrafiltrate were low, probably as a result of binding of the drug to the ultrafiltration system. The amount of this binding decreased in the order A greater than B greater than C. When using A and B, the first portion of ultrafiltrate obtained after centrifuging for 15 minutes must therefore be discarded. Using the Ultrafree filters this was not necessary. Plasma proteins were found to pass through the Ultrafree filter membrane when more than 0.7 ml ultrafiltrate was collected. Comparison of the ultrafiltration methods with equilibrium dialysis showed a good correlation: r = 0.96 (A), r = 0.97 (B) and r = 0.98 (C). However, in the carbamazepine plasma concentration range 2.5-10.0 micrograms/ml, significantly lower free fractions were measured with the Centriflo cones. The free fractions measured were: 0.28-0.30 (equilibrium dialysis), 0.17-0.18 (A), 0.22-0.24 (B) and 0.29-0.30 (C). The time required to collect the ultrafiltrate was 30 minutes (A and B) and 2 hours (C).
测试了三种简单的、市售的超滤系统[Centriflo CF25(A)、Centriflo CF50A(B)和Ultrafree(C)]在测定血浆中游离卡马西平浓度方面的实用性。在超滤物的第一部分中测得的卡马西平浓度较低,这可能是由于药物与超滤系统结合所致。这种结合量按A>B>C的顺序递减。因此,使用A和B时,离心15分钟后获得的超滤物的第一部分必须弃去。使用Ultrafree过滤器则无需如此。当收集超过0.7 ml超滤物时,发现血浆蛋白会穿过Ultrafree滤膜。超滤法与平衡透析法的比较显示出良好的相关性:r = 0.96(A),r = 0.97(B)和r = 0.98(C)。然而,在卡马西平血浆浓度范围为2.5 - 10.0微克/毫升时,使用Centriflo圆锥体测得的游离分数显著较低。测得的游离分数分别为:0.28 - 0.30(平衡透析),0.17 - 0.18(A),0.22 - 0.24(B)和0.29 - 0.30(C)。收集超滤物所需的时间为30分钟(A和B)和2小时(C)。