Barrows G H, Sisken J E, Allegra J C, Grasch S D
J Histochem Cytochem. 1984 Jul;32(7):741-6. doi: 10.1177/32.7.6736626.
The authors describe a system for the quantitation of fluorescence light output by individual cells using the signal obtained from a silicon intensifier target video camera. The video image is digitized to 4 bits (16 levels), and a 512 X 512 matrix is constructed and stored in 128K of video memory. Areas to be measured are user-specified by means of a light pen entry system. Recorded intensities are integrated under microprocessor control to provide a measure of total fluorescence in the selected areas. The distribution of light intensities per pixel over the 16-level gray scale as well as morphometric data are also obtainable. Linear response to transmission and fluorescence standards was verified. Reproducibility of the system was evaluated using fluorescent beads and glutaraldehyde-fixed chick red blood cells, which gave coefficients of variation comparable to those obtainable from other systems. Measurements of DNA per nucleus of human diploid fibroblasts using Hoechst dye 33258 yielded the two sharp peaks corresponding to the 2C and 4C values of DNA expected from such cells. We conclude that digital video measurement of fluorescence provides meaningful data and has considerable promise as a sensitive tool for the quantitation of fluorescence at the cellular level.
作者描述了一种使用从硅增强靶视频摄像机获得的信号对单个细胞发出的荧光光输出进行定量的系统。视频图像被数字化为4位(16级),构建一个512×512矩阵并存储在128K的视频存储器中。通过光笔输入系统由用户指定要测量的区域。在微处理器控制下对记录的强度进行积分,以提供所选区域总荧光的测量值。还可获得每个像素在16级灰度上的光强度分布以及形态测量数据。验证了对透射和荧光标准的线性响应。使用荧光珠和戊二醛固定的鸡红细胞评估了该系统的重现性,其变异系数与其他系统相当。使用Hoechst染料33258对人二倍体成纤维细胞核中的DNA进行测量,产生了与这类细胞预期的2C和4C DNA值相对应的两个尖峰。我们得出结论,荧光的数字视频测量提供了有意义的数据,作为一种在细胞水平定量荧光的灵敏工具具有很大的前景。