Lipetz L E
Vision Res. 1984;24(6):567-604.
A new method measures the spectral absorption of small volume samples having peak optical densities (Dp) too high (up to 50 and higher) to be measured by standard direct or indirect methods. This method specifically corrects for the effect on the microspectrophotometer (MSP) of wavelength-independent light which bypasses the sample. The method involves; (1) measuring by MSP the long wavelength cutoff of the dense samples and the absorption spectrum of thinned samples of the same pigment, (2) calculating Dp from these. The method's application is illustrated for the colored oil droplets in the retinal cones of the turtle, Emydoidea blandingii.
一种新方法可测量小体积样品的光谱吸收,这些样品的峰值光密度(Dp)过高(高达50及更高),无法通过标准的直接或间接方法进行测量。该方法专门校正了绕过样品的与波长无关的光对显微分光光度计(MSP)的影响。该方法包括:(1)通过MSP测量致密样品的长波长截止值以及相同色素稀释样品的吸收光谱,(2)据此计算Dp。以 Blandingii 锦龟视网膜视锥细胞中的有色油滴为例说明了该方法的应用。