Hanson W F, Berkley L W, Peterson M
Med Phys. 1980 Mar-Apr;7(2):145-6. doi: 10.1118/1.594677.
The effective energy of the x-ray beam from linear accelerators changes as a function of the position in the beam due to nonuniform filtration by the flattening filter. In this work, the transmittance through a water column was measured in good geometry and the beam quality characterized in units of HVL in water. Measurements were made on a variety of linear accelerators from 4 to 10 MV. The beam energy decreased with increasing distance from the central ray for all accelerators measured.
由于均整滤过器的非均匀滤过,直线加速器产生的X射线束的有效能量会随着束流中位置的变化而改变。在本研究中,在良好的几何条件下测量了穿过水柱的透射率,并以水中半价层(HVL)为单位对射束质量进行了表征。对4至10MV的各种直线加速器进行了测量。对于所有测量的加速器,射束能量随着与中心射线距离的增加而降低。