Paynter D, Weston S J, Cosgrove V P, Evans J A, Thwaites D I
St James Institute of Oncology The Leeds Teaching Hospitals NHS Trust, Medical Physics, Leeds LS9 7TF, United Kingdom.
LIGHT Institute University of Leeds Leeds LS2 9JT, Division of Medical Physics, Leeds, United Kingdom.
Med Phys. 2014 May;41(5):052103. doi: 10.1118/1.4871615.
Flattening filter free (FFF) linear accelerators can increase treatment efficiency and plan quality. There are multiple methods of defining a FFF beam. The Elekta control system supports tuning of the delivered FFF beam energy to enable matching of the percentage depth-dose (PDD) of the flattened beam at 10 cm depth. This is compared to FFF beams where the linac control parameters are identical to those for the flattened beam. All beams were delivered on an Elekta Synergy accelerator with an Agility multi-leaf collimator installed and compared to the standard, flattened beam. The aim of this study is to compare "matched" FFF beams to both "unmatched" FFF beams and flattened beams to determine the benefits of matching beams.
For the three modes of operation 6 MV flattened, 6 MV matched FFF, 6 MV unmatched FFF, 10 MV flattened, 10 MV matched FFF, and 10 MV unmatched FFF beam profiles were obtained using a plotting tank and were measured in steps of 0.1 mm in the penumbral region. Beam penumbra was defined as the distance between the 80% and 20% of the normalized dose when the inflection points of the unflattened and flattened profiles were normalized with the central axis dose of the flattened field set as 100%. PDD data was obtained at field sizes ranging from 3 cm × 3 cm to 40 cm × 40 cm. Radiation protection measurements were additionally performed to determine the head leakage and environmental monitoring through the maze and primary barriers.
No significant change is made to the beam penumbra for FFF beams with and without PDD matching, the maximum change in penumbra for a 10 cm × 10 cm field was within the experimental error of the study. The changes in the profile shape with increasing field size are most significant for the matched FFF beam, and both FFF beams showed less profile shape variation with increasing depth when compared to flattened beams, due to consistency in beam energy spectra across the radiation field. The PDDs of the FFF beams showed less variation with field size, the d(max) value was deeper for the matched FFF beam than the FFF beam and deeper than the flattened beam for field sizes greater than 5 cm × 5 cm. The head leakage when using the machine in FFF mode is less than half that for a flattened beam, but comparable for both FFF modes. The radiation protection dose-rate measurements show an increase of instantaneous dose-rates when operating the machines in FFF mode but that increase is less than the ratio of MU/min produced by the machine.
The matching of a FFF beam to a flattened beam at a depth of 10 cm in water by increasing the FFF beam energy does not reduce any of the reported benefits of FFF beams. Conversely, there are a number of potential benefits resulting from matching the FFF beam; the depth of maximum dose is deeper, the out of field dose is potentially reduced, and the beam quality and penetration more closely resembles the flattened beams currently used in clinical practice, making dose distributions in water more alike. Highlighted in this work is the fact that some conventional specifications and methods for measurement of beam parameters such as penumbra are not relevant and further work is required to address this situation with respect to "matched" FFF beams and to determine methods of measurement that are not reliant on an associated flattened beam.
无均整器(FFF)直线加速器可提高治疗效率和计划质量。定义FFF射束有多种方法。医科达控制系统支持对输出的FFF射束能量进行调谐,以使均整射束在10 cm深度处的百分深度剂量(PDD)相匹配。将其与直线加速器控制参数与均整射束相同的FFF射束进行比较。所有射束均在安装了敏捷多叶准直器的医科达Synergy加速器上输出,并与标准均整射束进行比较。本研究的目的是将“匹配的”FFF射束与“不匹配的”FFF射束以及均整射束进行比较,以确定射束匹配的益处。
对于6 MV均整、6 MV匹配FFF、6 MV不匹配FFF、10 MV均整、10 MV匹配FFF和10 MV不匹配FFF这三种运行模式,使用绘图水箱获取射束轮廓,并在半影区域以0.1 mm的步长进行测量。射束半影定义为当未均整和均整轮廓的拐点以均整野中心轴剂量设为100%进行归一化时,归一化剂量80%和20%之间的距离。在3 cm×3 cm至40 cm×40 cm的射野尺寸范围内获取PDD数据。另外进行了辐射防护测量,以确定机头泄漏以及通过迷宫和主防护屏蔽进行的环境监测。
对于有和没有PDD匹配的FFF射束,射束半影没有显著变化,对于10 cm×10 cm射野,半影的最大变化在研究的实验误差范围内。随着射野尺寸增加,轮廓形状的变化对于匹配的FFF射束最为显著,并且与均整射束相比,两种FFF射束随着深度增加轮廓形状变化更小,这是由于辐射野内射束能谱的一致性。FFF射束的PDD随射野尺寸变化较小,对于射野尺寸大于5 cm×5 cm的情况,匹配的FFF射束的d(max)值比FFF射束更深且比均整射束更深。使用机器处于FFF模式时的机头泄漏小于均整射束的一半,但两种FFF模式相当。辐射防护剂量率测量表明,机器在FFF模式运行时瞬时剂量率增加,但增加幅度小于机器产生的MU/min的比值。
通过增加FFF射束能量使FFF射束在水中10 cm深度处与均整射束匹配,不会降低FFF射束已报道的任何益处。相反,匹配FFF射束会带来许多潜在益处;最大剂量深度更深,射野外剂量可能降低,并且射束质量和穿透更接近目前临床实践中使用的均整射束,使水中剂量分布更相似。这项工作突出了这样一个事实,即一些用于测量射束参数(如半影)的传统规范和方法并不适用,需要进一步开展工作来解决关于“匹配的 ”FFF射束的这种情况,并确定不依赖相关均整射束的测量方法。