Suppr超能文献

使用自动离散分析仪对血清和脑脊液中的白蛋白和免疫球蛋白G进行快速免疫比浊测定。

Rapid immunoturbidimetric assay of albumin and immunoglobulin G in serum and cerebrospinal fluid with an automatic discrete analyser.

作者信息

Kamp H H, Luderer T K, Muller H J, Sopjes-Kruk A

出版信息

Clin Chim Acta. 1981 Aug 10;114(2-3):195-205. doi: 10.1016/0009-8981(81)90392-2.

Abstract

A method is presented to determine albumin and immunoglobulin G (IgG) in serum and cerebrospinal fluid (CSF) with an automatic discrete analyser. Concentrations of albumin and IgG in serum and CSF are compared with values obtained by radial immunodiffusion (RID). Ratios of albumin in CSF over albumin in serum and the so-called IgG index are calculated. Ratios obtained with the discrete analyser and with the RID are compared and discussed with respect to clinical diagnosis.

摘要

介绍了一种使用自动离散分析仪测定血清和脑脊液(CSF)中白蛋白和免疫球蛋白G(IgG)的方法。将血清和脑脊液中白蛋白和IgG的浓度与通过放射免疫扩散(RID)获得的值进行比较。计算脑脊液中白蛋白与血清中白蛋白的比率以及所谓的IgG指数。就临床诊断而言,比较并讨论了用离散分析仪和RID获得的比率。