Suppr超能文献

麻醉气体对皮肤表面氧气和二氧化碳传感器的干扰。

Interference of anesthetic gases at skin surface sensors for oxygen and carbon dioxide.

作者信息

Eberhard P, Mindt W

出版信息

Crit Care Med. 1981 Oct;9(10):717-20. doi: 10.1097/00003246-198110000-00010.

Abstract

Several variables may account for the response of electrochemical skin surface PO2 sensors to anesthetic gases: cathode material and size, pH of the electrolyte and membrane material. These variables cannot be chosen arbitrarily and their influence has been tested with two types of sensors. In one type (LSC), a large size cathode (mm range) and a membrane with low permeability for oxygen such as mono-axially oriented polyethylene is used. The other type (MC) contains one or more microcathodes (micron range) and a membrane which is highly permeable for oxygen such as Teflon PTFE. With the LSC sensor, the N2O interference current is smaller than 5% of the air current when the sensor is polarized at --600 mV. The interference current with 2% halothane is smaller than 3% of the air current. With the MC sensor, the N2O interference may be up to 40% of the current in air when the sensor is polarized at --800 mV. The magnitude of this interference depends considerably on the silver deposition on the platinum cathode. At --600 mV the N2O interference is negligible. However, at this polarization voltage, the sensor is not operated within the limiting current plateau of oxygen. The interference current with 2% halothane may be up to 30% of the current in air. With both types of sensors there was no measurable interference by 2% enflurane. The authors conclude that to reduce the interference of anesthetic gases at skin surface sensors for oxygen to a reasonable level, it is necessary to use a membrane with low permeability for oxygen and a polarization voltage of approximately --600 mV. These two conditions can be fulfilled optimally only with a sensor design in which a large size cathode is used. At Stowe-Severinghaus type skin surface sensors for PCO2, there is no measurable interference by N2O, halothane or enflurane.

摘要

有几个变量可能会影响电化学皮肤表面PO2传感器对麻醉气体的响应:阴极材料和尺寸、电解质的pH值以及膜材料。这些变量不能随意选择,并且已经用两种类型的传感器测试了它们的影响。在一种类型(LSC)中,使用大尺寸阴极(毫米范围)和对氧气渗透性低的膜,如单轴取向聚乙烯。另一种类型(MC)包含一个或多个微阴极(微米范围)和对氧气高渗透性的膜,如聚四氟乙烯(Teflon PTFE)。对于LSC传感器,当传感器在-600 mV极化时,N2O干扰电流小于空气电流的5%。2%氟烷的干扰电流小于空气电流的3%。对于MC传感器,当传感器在-800 mV极化时,N2O干扰可能高达空气中电流的40%。这种干扰的大小在很大程度上取决于铂阴极上的银沉积。在-600 mV时,N2O干扰可以忽略不计。然而,在这个极化电压下,传感器不在氧气的极限电流平台内运行。2%氟烷的干扰电流可能高达空气中电流的30%。对于这两种类型的传感器,2%恩氟烷均无明显干扰。作者得出结论,为了将皮肤表面氧气传感器上麻醉气体的干扰降低到合理水平,有必要使用对氧气渗透性低的膜和大约-600 mV的极化电压。只有采用大尺寸阴极的传感器设计才能最佳地满足这两个条件。对于斯托韦-西弗林豪斯型皮肤表面PCO2传感器,N2O、氟烷或恩氟烷均无明显干扰。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验