Haas R G, Mushel S
Am J Clin Pathol. 1982 Feb;77(2):216-9. doi: 10.1093/ajcp/77.2.216.
The presence of hemoglobin in serum or plasma causes an increase in the apparent calcium concentration when this determination is performed on the DuPont Automatic Clinical Analyzer (aca). A simple pack modification is described using ethylene glycol-bis-(2-aminoethylether)-tetracetic acid (EGTA) which permits estimation of the positive interference due to hemoglobin color. A correction factor is applied to enable the accurate measurement of calcium in hemolyzed samples.