Rosen I, Smith A, Lane R, Kelsey C, Lake D, Hogstrom K, Somers J, Helland J, Kittell R, Amols H, Bradbury J, Richman C
Med Phys. 1978 Mar-Apr;5(2):120-3. doi: 10.1118/1.594418.
An automated data-acquisition and analysis system has been developed for dosimetry measurements on the pion therapy beam at the Clinton P. Anderson Meson Physics Facility Biomedical Channel in Los Alamos using a PDP-11/45 computer and CAMAC interface. Initialization, test, and monitor programs allow the user to set the physical limits of scanner travel, test the data lines, calibrate the analog signals for the scanner position, and monitor the analog versus digital values of the scanner position during operation. Data-acquisition programs scan beams in one, two, and three dimensions. Many options are available to the user in selecting the scan parameters and in changing some of these parameters during scanning. Data-analysis programs provide reproduction of stored data, comparison of linear scans, beam profiles along any line of a planar or volume scan, and isodose distributions from any planar scan or from any planar scan or from any plane of a volume scan. Other programs summarize stored data files and search for specific data according to the user's instruments.
已开发出一种自动数据采集与分析系统,用于在洛斯阿拉莫斯的克林顿·P·安德森介子物理设施生物医学通道的π介子治疗束上进行剂量测定,该系统使用PDP - 11/45计算机和CAMAC接口。初始化、测试和监控程序允许用户设置扫描仪行程的物理极限、测试数据线、校准扫描仪位置的模拟信号,并在操作过程中监控扫描仪位置的模拟值与数字值。数据采集程序可在一维、二维和三维中扫描束流。用户在选择扫描参数以及在扫描过程中更改其中一些参数时有多种选项。数据分析程序可再现存储的数据、比较线性扫描、给出平面或体积扫描中任意一条线的束流剖面,以及给出任意平面扫描或体积扫描中任意平面的等剂量分布。其他程序可总结存储的数据文件,并根据用户的仪器搜索特定数据。