Suppr超能文献

高效液相色谱法中使用铜电极安培检测氨基酸。

Amperometric detection of amino acids in high-performance liquid chromatography with a copper electrode.

作者信息

Kok W T, Brinkman U A, Frei R W

出版信息

J Chromatogr. 1983 Jan 28;256(1):17-26. doi: 10.1016/s0021-9673(01)88208-x.

Abstract

A copper electrode has been used as an amperometric detector for amino acids in high-performance liquid chromatography. The amino acids are separated in a reversed-phase system, using silica-based and polymer-type column materials. Neutral or alkaline buffer solutions of phosphate and carbonate can be used as mobile phases. Borate buffers are less suitable. The detection method is characterized by a high linear dynamic range, good reproducibility, the absence of electrode poisoning and a sensitivity comparable to that of UV absorption methods after derivatization of the amino acids. Detection limits with conventional-scale columns are in the range 10-100 pmoles. A reduction in the flow-rate in the flow-through cell improves the sensitivity for amino acids that give relatively low signals, such as proline. Therefore, the use of microbore columns is especially advantageous for these compounds. The absolute detection limits decrease by about one order of magnitude on changing to a miniaturized system.

摘要

铜电极已被用作高效液相色谱中氨基酸的安培检测器。氨基酸在反相系统中分离,使用基于硅胶和聚合物类型的柱材料。磷酸盐和碳酸盐的中性或碱性缓冲溶液可用作流动相。硼酸盐缓冲液不太适合。该检测方法的特点是线性动态范围宽、重现性好、不存在电极中毒现象,并且在氨基酸衍生化后灵敏度与紫外吸收法相当。使用常规尺寸色谱柱时的检测限在10 - 100皮摩尔范围内。降低流通池中的流速可提高对信号相对较低的氨基酸(如脯氨酸)的灵敏度。因此,使用微径柱对这些化合物特别有利。改用小型化系统后,绝对检测限降低约一个数量级。