Lott J A, Hartzell R K, Longberry J
Am J Med Technol. 1983 Jan;49(1):43-8.
At present, the quality control of platelet counting by semi-automated and automated methods does not meet ideal standards. Controls prepared from human or animal platelets have limited stability, and some synthetic platelet controls that are available do not have the size distribution of fresh platelets. The platelet control materials described here are wholly synthetic; however, their particle size distribution is like that of normal human platelets, and the dispersing medium has the viscosity and surface tension of plasma. Two types of products are described. The first type are dilutions of the synthetic platelets which are handled like 3000-fold dilutions of platelet-rich plasma and are intended for direct use on instruments like the Coulter ZBI. The two dilution levels gave counts of about 50,000 and 200,000/microL on the Coulter ZBI and were found to be stable for at least 30 days at - 20C, 4C, and 37C, and at least eight months at 25C. The second type of product is handled like whole blood and is intended for direct use on instruments like the Coulter Model S-Plus. This product gave counts of about 200,000/microL and was found to be stable for at least 120 days at - 20C, 4C, 25C, and 37C. Freezing at - 20C produced some aggregates that dispersed after thawing and standing for several days prior to testing.
目前,半自动和自动方法进行血小板计数的质量控制未达到理想标准。由人或动物血小板制备的对照品稳定性有限,现有的一些合成血小板对照品没有新鲜血小板的大小分布。这里描述的血小板对照材料完全是合成的;然而,它们的粒径分布与正常人血小板相似,分散介质具有血浆的粘度和表面张力。描述了两种类型的产品。第一种是合成血小板的稀释液,其处理方式类似于富血小板血浆的3000倍稀释液,旨在直接用于库尔特ZBI等仪器。这两种稀释水平在库尔特ZBI上的计数约为50,000和200,000/微升,发现在-20°C、4°C和37°C下至少稳定30天,在25°C下至少稳定八个月。第二种产品的处理方式类似于全血,旨在直接用于库尔特S-Plus型等仪器。该产品的计数约为200,000/微升,发现在-20°C、4°C、25°C和37°C下至少稳定120天。在-20°C冷冻会产生一些聚集体,解冻后并在测试前静置几天会分散。