[家庭护理吸入装置中的微生物]

[Microorganisms in home-care inhalation devices].

作者信息

Exner M, Vogel F, Rost H D

出版信息

Dtsch Med Wochenschr. 1983 Jan 7;108(1):12-7. doi: 10.1055/s-2008-1069492.

Abstract

Contamination of inhalation equipment used at home, cleaning procedures, and the occurrence of potentially pathogenic micro-organisms in mouthwash water was studied in 55 patients. Fungi in the oropharyngeal space were found in 58%, Staphylococcus aureus in 29%, Enterobacteriaceae in 47% and Pseudomonas aeruginosa in 11%. Contamination with potentially pathogenic micro-organisms occurred in 64% of inhalation equipment and inhalation solutions. In a total of 151 samples from the equipment Klebsiella was present in 14%, Enterobacter in 10%, Serratia and Acinetobacter in 11% each, Pseudomonas in 5% and Staphylococcus aureus in 4.5%. In three cases the same serotype of Ps. aeruginosa was present in the oropharyngeal cavity and inhalation equipment. In most cases the cleansing of the inhalation equipment had been inadequate. Only 36% of those questioned cleaned the equipment immediately before use, only 4 of 20 stated that they washed their hands before use. Subjective assessment of the value of inhalation treatment was either "very good" or "good" in 80%.

摘要

对55例患者的家用吸入设备污染情况、清洁程序以及漱口水中潜在致病微生物的出现情况进行了研究。口咽部位检出真菌的比例为58%,金黄色葡萄球菌为29%,肠杆菌科为47%,铜绿假单胞菌为11%。64%的吸入设备和吸入溶液存在潜在致病微生物污染。在设备的151份样本中,克雷伯菌属占14%,肠杆菌属占10%,沙雷菌属和不动杆菌属各占11%,假单胞菌属占5%,金黄色葡萄球菌占4.5%。3例患者的口咽腔和吸入设备中存在相同血清型的铜绿假单胞菌。在大多数情况下,吸入设备的清洁不充分。只有36%的受访者在使用前立即清洁设备,20人中只有4人表示在使用前洗手。80%的患者对吸入治疗效果的主观评价为“非常好”或“好”。