Pacemaker-mediated tachycardias: a rapid bedside technique for induction and observation.

作者信息

Littleford P O, Curry R C, Schwartz K M, Pepine C J

出版信息

Am J Cardiol. 1983 Aug;52(3):287-91. doi: 10.1016/0002-9149(83)90124-8.

Abstract

The DDD pacemaker allows sensing and pacing in both chambers and can, therefore, maintain atrioventricular synchrony. However, this pacemaker creates an additional anterograde conduction system between the atrium and ventricle and, in the presence of ventriculoatrial (VA) conduction, the possibility of pacemaker-mediated tachycardia exists. A simple bedside technique that does not require catheterization or expensive equipment was used to detect VA conduction. Just after DDD pacer implantation, an attempt was made to detect VA conduction in 31 patients. Ambulatory monitoring (Holter) was done for 24 hours after implantation and at 2 to 4 and 6 to 8 weeks after implantation to detect pacemaker-mediated tachycardia. Attempts to induce pacemaker-mediated tachycardia were made using a special programmable external stimulator at follow-up after implantation. It was found that (1) all the 17 patients with detectable VA conduction had pacemaker-mediated tachycardia when the atrial refractory period was less than the VA conduction time, (2) pacemaker-mediated tachycardias were not inducible or detected spontaneously when atrial refractory period was equal to or greater than VA conduction time + 50 ms, (3) VA conduction was not detectable in 9 of the 17 patients at a later visit. Six of these 9 received antiarrhythmic therapy or had developed congestive heart failure. VA conduction has important implications in patients with DDD pacemakers and can be readily evaluated at the bedside.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验