Riquin D, Fankhauser F, Lörtscher H
Int Ophthalmol. 1983 Jun;6(3):191-200. doi: 10.1007/BF00141128.
Two new contact glasses, one for irradiation of the iris and the other for irradiation of pathological structures within the pupillary and retropupillary regions are described. First order geometric considerations using the Young-Weierstrass imaging theorem indicate nearly aberration-free, stigmatic focussing, even for large numerical apertures. For surgical tasks utilizing high-power lasers, excellent focussing of strongly convergent ray-bundles which lead to small focus diameter and large diameters while passing through the pre-focal optical media are seen to be essential. Thus the efficiency of the laser-surgical treatment, its reliability, as well as security and harmlessness within the pre-focal media are increased.
介绍了两种新型隐形眼镜,一种用于照射虹膜,另一种用于照射瞳孔和瞳孔后区域内的病理结构。利用杨氏-魏尔斯特拉斯成像定理进行的一阶几何考量表明,即使对于大数值孔径,也几乎无像差、无像散聚焦。对于使用高功率激光的手术任务而言,能实现强会聚光束的出色聚焦,这在光束穿过焦前光学介质时会形成小的焦斑直径和大的光束直径,而这一点至关重要。因此,激光手术治疗的效率、可靠性以及焦前介质内的安全性和无害性都得以提高。